Advanced Programming
5-7
January 2008
The Trim Submenu
At the following prompt, the XDP analog output of the meter is
driven at exactly 4 mA.
At the following prompt, the XDP analog output of the meter is
driven at exactly 20 mA.
To continue programming the setup menu, proceed to the appropriate
section for instructions. Otherwise, press
[ESC]
four times to return to
live measurement mode.
4-20mA-Out
Use the [
S
] and [
T
] keys to
select
[Trim]
and press
[ENTER]
.
4-20_Out_Range
Test
Trim
4-20mA-Out Trim
Press the [
S
] and [
T
] key until
your output device reads
exactly 4 mA. When done,
press
[ENTER]
.
4mA-Trim
s
x
t
4-20mA-Out Tr..Trim
Press the [
S
] and [
T
] key until
your output device reads
exactly 20 mA. When done,
press
[ENTER]
.
20mA-Trim
s
x
t
4-20mA-Out
Press
[ESC]
to return to the
setup menu.
4-20_Out_Range
Test
Trim