January 2008
5-8
Advanced Programming
Configuring the Low and
Low Low Alarms
To configure the XDP Low and Low Low alarms, access the setup
menu as described on page 5-4 and complete the following steps:
Note:
The other six alarm relays listed in the alarm setup menu are
discussed in the next section.
The Low Alarm
At the next prompt, enter the percentage of hydrogen in air that
should trip the low alarm (typically 85.00%).
Range
Cal
Setup
Use the [
S
] and [
T
] keys to
select
[Alarms]
and press
[ENTER]
.
4-20mA-Out
Alarms
Alarms
Use the [
S
] and [
T
] keys to
select
[Low-Alarm]
and press
[ENTER]
.
Low-Alarm
Low-Low-Alarm
Normal-Alarm
Alarms Low-Alarm
Use the [
S
] and [
T
] keys to
select
[Trip-Point]
and press
[ENTER]
.
Trip-Point
Test
Fail-Safe
Alarms..Trip-Point
Use [
W
] and [
X
] to position the
cursor under the desired char-
acter. Then, use [
S
] and [
T
] to
increment the value. Press
[ENTER]
when done.
Trip-Point
xxx.xx
s
Alarms Low-Alarm
Use the [
S
] and [
T
] keys to
select
[Test]
and press
[ENTER]
.
Trip-Point
Test
Fail-Safe
Summary of Contents for XDP-H2
Page 3: ...Chapter 1 ...
Page 9: ...Chapter 2 ...
Page 20: ...Chapter 3 ...
Page 28: ...Chapter 4 ...
Page 55: ...Chapter 5 ...
Page 76: ...Chapter 6 ...
Page 80: ...Appendix A ...
Page 81: ...CE Mark Compliance Introduction A 1 EMC Compliance A 1 LVD Compliance A 1 ...
Page 83: ...Appendix B ...
Page 87: ...Appendix C ...