January 2008
B-2
Additional Wiring Connections
XDP Alarm Relays
In addition to the alarm relays discussed in Chapter 2, Installation,
there are four supplemental alarm relays at terminal block J7.
•
Gas Curve Relay 1
•
Gas Curve Relay 2
•
Process/Cal Relay
•
Zero/Span Relay
Note:
Contact GE Sensing for information on the use and functions
of these alarms.
To wire the supplemental alarms, refer to Figure 2-2 on page 2-10 and
Table B-1 above. Then, complete the following steps:
Note:
A
fail-safe
alarm is wired to the normally-closed (NC)
contacts, while a
non-fail-safe
alarm is wired to the normally-
open (NO) contacts.
1.
For
non-fail-safe
operation, connect pin #1 to the alarm input.
2.
Connect pin #2 to the alarm device return.
3.
For
fail-safe
operation, connect pin #3 to the alarm input
Table B-1: Terminal Block J7 Connections
Pin #
Alarm
Terminal
1
Gas Curve Relay 1
NC
2
C
3
NO
4
Gas Curve Relay 2
NC
5
C
6
NO
7
Process/Cal Relay
NC
8
C
9
NO
10
Zero/Span Relay
NC
11
C
12
NO