January 2008
4-24
Setup and Calibration
Performing a Span
Calibration (cont.)
At the next prompt, the new span drift value and an indication that the
meter passed or failed the calibration attempt are displayed.
Note:
If the calibration “
Failed”
, repeat the calibration procedure.
If another failure occurs, contact the factory for assistance.
Press
[ESC]
twice to return to the main menu. Then, press
[ESC]
again
to exit the
User Program
, or proceed to the appropriate section of this
chapter to continue programming the meter.
Viewing and Clearing the
Cal Drift
The XDP internally monitors the amount of drift from the factory
calibration, and the operator may view the drift values and clear them,
if desired. To accomplish this, access the main menu of the
User
Program
(see page 3-4) and proceed as follows:
Gas-Curve: Man..Cal
If the calibration
“Passed”
,
use [
W
] and [
X
] to highlight
[Finish]
. Then, press the
[ENTER]
key.
Span Drift
x.xx mA Passed
<Back
Finish
Gas-Curve: Manual..
Press
[ESC]
three times to
return to the
Cal Menu
.
Zero
Span
Cal
Opt
Disp
If necessary, use the [
W
] and
[
X
] keys to select
[Cal]
and
press
[ENTER]
.
Cal
Opt
Disp
Press
[ENTER]
.
1:Cal
Range
Cal
Use the [
W
] and [
X
] keys to
select
[Cal]
and press
[ENTER]
.
Slct Cal Menu Item
Summary of Contents for XDP-H2
Page 3: ...Chapter 1 ...
Page 9: ...Chapter 2 ...
Page 20: ...Chapter 3 ...
Page 28: ...Chapter 4 ...
Page 55: ...Chapter 5 ...
Page 76: ...Chapter 6 ...
Page 80: ...Appendix A ...
Page 81: ...CE Mark Compliance Introduction A 1 EMC Compliance A 1 LVD Compliance A 1 ...
Page 83: ...Appendix B ...
Page 87: ...Appendix C ...