Section 3-4 - Completing the Installation
Section 3-4
Completing the Installation
3-4-1
System Specifications
3-4-1-1
Physical Dimensions of Voluson® 730
The physical dimensions of the Voluson® 730 unit are summarized in
Table 3-5
.
Table 3-6
lists the size of Voluson® 730, with monitor and without on-board peripherals.
3-4-1-2
Weight without Monitor and Peripherals
3-4-1-3
Acoustic Noise Output
max. 57dB(A)
3-4-2
Electrical Specifications
Power Consumption nominal 1010 VA including all options.
Mains outlets: Mains socket ST1, ST2, ST3, ST4, ST5 for accessories.
All mains outlets are co-switched by the unit’s mains switch via built-in isolation transformer.
Output voltage for: ST1 - ST5: 115V or 230V.
Table 3-5
Physical Dimensions of Voluson® 730
Height
Width
Depth
145 cm / 57.1 inches
68 cm / 26.8 inches
100 cm / 39.4 inches
Table 3-6
Weight of Voluson® 730 With Monitor and Without Other Peripherals
Model
Weight [kg]
Weight [lbs.]
Voluson® 730
136
300
Table 3-7
Electrical Specifications for Voluson® 730
Voltage
Tolerances
Current
Frequency
100 VAC
±10%
10.10 A
50, 60 Hz (±2%)
115 VAC
±10%
8.80 A
50, 60 Hz (±2%)
130 VAC
±10%
7.80 A
50, 60 Hz (±2%)
230 VAC
±10%
4.40 A
50, 60 Hz (±2%)
240 VAC
±10%
4.20 A
50, 60 Hz (±2%)
CAUTION
Modification of voltage setting only by an authorized service person! Output power: 350VA per
mains outlet, max. power of all connected accessories must not exceed 350VA.