Chapter 5 Components and Functions (Theory)
5-17
5-2-6
Description of Hardware Options
5-2-6-1
CW - Continuous Wave Doppler
CW Doppler mode provides real time spectral analysis of CW Doppler signals. This information
describes the Doppler shifted signal from the moving reflectors in the CW Doppler beam. CW Doppler
can be referenced through a small pencil probe or phased array scan head, but it can also be used in
conjunction with a 2D image which has an M-line superimposed on the 2D image indicating the position
of the Doppler sample volume. For through-the-beamformer CW, this beam is steerable by the operator,
and is done by adjusting the location of the M-line. The CW Doppler beam, or M-mode line, can be
steered allowing interrogation along an operator-selected line within the image.
This option can be upgraded by implementing the CW-Dopplerboard (CRW, old version CWD).
5-2-6-2
ECG Digital Module
The ECG-preamplifier is used for acquiring an ECG-signal to be displayed with the ultrasound image.
This optional peripheral serves for gaining an ECG-signal to mark the systolic and end diastolic
moments in M-Mode and Doppler evaluations.
The ECG-preamplifier must not be used for ECG-diagnostics. It is not intended for use as a cardiac
monitor and must not be used for an intra-operative application on the heart.
1.) MAN3 (external version)
2.) MAN6 (internal, digital version)
5-2-6-3
Scan freeze Footswitch
Table 5-3
Hardware Options
HW-Options
Description
1
CW-Doppler
5-2-6-1 CW - Continuous Wave Doppler
2
ECG Digital Module
5-2-6-2 ECG Digital Module
3
Scan freeze Footswitch
5-2-6-3 Scan freeze Footswitch
Figure 5-6 Foot-switch Connector on CPN
Footswitch connected to Power Supply-Box (below Main Electronic-Box)