Chapter 4 Functional Checks
4-31
4-4-13
ECG Check Out
Connect the ECG-preamplifier MAN and check:
Section 4-5
Software Configuration Checks
Touch the UTILITIES key, then touch SYSTEM SETUP. The System Setup desktop offers different
pages to check:
Touch the UTILITIES key, then touch MEASURE SETUP.
The Measurement Setup desktop offers different pages to check:
Table 4-10 ECG-preamplifier Check
Step
Task
Expected Result(s)
1
Connect the ECG at the Connector on the rear panel of the
scanner. Touch the UTILITIES key, then the ECG key on
the Touch Panel.
It will display a curve along the bottom edge of the image
sector
Table 4-11 Software Configuration (System Setup) Checks
Step
Task
Expected Result(s)
1
General: Check Date and Time setting
Date and Time are correct
2
General: Check that Location (Clinic Name) is correct
Location Name is correct
3
General: Check Language settings
desired Language is displayed
4
User Setting: Check all the User Settings
settings assigned as desired by the customer
5
Peripherals: Check assignment of Printer Keys
Print A and Print B keys are assigned as desired by the
customer
6
Peripherals: Check assignment of Foot Switch
Foot Switch left and right are assigned as desired by the
customer
7
Peripherals: Check Save Destination assignment
assigned as desired by the customer
8
Peripherals: Check 2D Save key assignment
assigned as desired by the customer
9
Peripherals: Check Remote Print A, Remote Print B,
Report Printer and DICOM Print Job assignment
settings assigned as desired by the customer
10
Option: Check that all of the options are set up correct
all authorized functions are enabled
Table 4-12 Measurement Setup Checks
Step
Task
Expected Result(s)
1
General: Check “2D Circumference and Area Method”
setting assigned as desired by the customer
2
General: Check “Show Caliper Information Text”
setting assigned as desired by the customer
3
General: Check “Caliper & Measure-Result Font Size”
setting assigned as desired by the customer
4
General: Check “Doppler Manual Trace Mode”
setting assigned as desired by the customer