GE M
EDICAL
S
YSTEMS
- K
RETZTECHNIK
U
LTRASOUND
D
IRECTION
105844, R
EVISION
1
V
OLUSON
® 730 S
ERVICE
M
ANUAL
3-16
Section 3-5 - Printer Installation
3-5-3
Installing Digital Color Printer (Sony UP-D2600S or Mitsubishi CP770DW)
1.) Power off/Shutdown the system as described in:
Section 3-4-4 on page 3-10
.
2.) Remove the rear cover plate as described in:
Section 3-3-1-2 on page 3-6
.
3.) Remove SCSI-Adapter.
4.) Feed the SCSI-cables through the holes of the system housing to the front side where the printer
will be placed and connect them to the printer.
5.) Feed the Power cable to the backside and connect it on the Voluson® 730 power supply.
NOTE:
For connection schemes refer to Chapter 20 in the Basic User Manual of Voluson® 730.
6.) Reinstall the rear cover plate.
NOTE:
“Mouse functions” can be performed by using the trackball for moving the cursor.
“Normal select” (Click) by pressing the left or right trackball keys and for “opening a context
menu” press the upper trackball key.
7.) Turn ON the printer.
8.) Switch ON the power of the system and wait till the system has booted.
NOTE:
The power switch of the printer has to be in ON position before starting the system!
Leave the power switch of the printer always in the ON position.
9.) Follow the printer installation steps as described in
Section 3-5-4 "Printer Installation manually" on
page 3-17
.
NOTICE
If only one single SCSI-cable is connected between Main Module and GEM (Drive Module):
Disconnect the SCSI connector from the main module (PC-part) and connect it to the SCSI-printer.
Connect the additional SCSI-cable (included in PZP50 connection set) from the main module to the
SCSI-printer.
NOTICE
Do not insert more SCSI-cables than required; the SCSI-cable length should be as short as possible!