Section 4-4 - Functional Checks
4-4-12-1
Formatting Media
(cont’d)
By selecting CD-RW the unit displays the “Erase CD-RW” window as shown in
Figure 4-18
.
1.) Select the “Erase Mode” and click the OK button to start the format process.
2.) During erasing the Ultrasound system displays following message.
3.) Touch the EXIT key on the Touch Panel to return to the Scan Mode.
NOTE:
It is possible to adjust the Write Speed of the CD-Writer.
At older software versions, also the Erase Mode can be preset in the “Settings” window.
1.) Click the SETTINGS icon on the left side of the Sonoview screen.
2.) Select
CD RECORDER
from the tool bar.
3.) Choose the “Write Speed” (and the “Erase Mode” at systems with older software versions)
from the corresponding pop-up menu.
NOTICE
It is highly recommended to use the complete erase mode, to avoid problems with the CD-RW!
Figure 4-18 Erase CD-RW Window
Figure 4-19 Erasing disk
Figure 4-20 Settings of the CD-Recorder