3. Operating in Phased Array Mode
PHASOR XS Operating Manual
Page 35
• PSA/—Minimum soundpath distance of all
beams in scan that are currently captured by
Gate A
• PPA/—Minimum projection distance of all
beams in scan that are currently captured by
Gate A
• PDA/—Minimum material depth of all beams
in scan that are currently captured by Gate A
• PZA/—Minimum uncorrected depth of all
beams in scan that are currently captured by
Gate A
• PSB/—Minimum soundpath distance of all
beams in scan that are currently captured by
Gate B
• PPB/—Minimum projection distance of all
beams in scan that are currently captured by
Gate B
• PDB^—Minimum material depth of all beams in
scan that are currently captured by Gate B
• PZB/—Minimum uncorrected depth of all
beams in scan that are currently captured by
Gate B
• SA^—Uncorrected Sound-Path distance or
duration represented by the highest echo to cross
A-Gate in the beam selected by the beam cursor.
• SA/—Uncorrected Sound-Path distance or
duration represented by the flank echo to cross
A-Gate in the beam selected by the beam cursor.
• SBA—Sound-Path distance or duration
between the highest echo (in the beam selected)
to cross B-Gate and the highest echo (in the
beam selected) to cross A-Gate.
• PA^—Projection distance from the beam index
point (BIP) to the reflector represented by the
highest echo in A-Gate (in the beam selected
by the beam cursor). (see
Figure 3-3
)
• PA/—Projection distance from the beam index point
(BIP) to the reflector represented by the flank echo
in A-Gate (in the beam selected by the beam cur-
sor). (see
Figure 3-3
)
The IF Gate provides a reference point from which the
other gates are positioned, and provides a reference
from which the displayed data can be positioned. No
results are based on IF-Gate triggering.
FIGURE 3-3—Various measured values available when a probe
wedge is used.
Summary of Contents for Phasor 16/16
Page 6: ...Contents Page 4 PHASOR XS Operating Manual ...
Page 16: ...1 Phased Array Operation Keypad Menus and Displays Page 14 PHASOR XS Operating Manual ...
Page 24: ...2 Setting Up for Phased Array Measurement Page 22 PHASOR XS Operating Manual ...
Page 86: ...5 Making Conventional Measurements Page 84 PHASOR XS Operating Manual ...
Page 92: ...6 Storing Data Sets and Generating Reports Page 90 PHASOR XS Operating Manual ...
Page 96: ...7 Technical Specifications Page 94 PHASOR XS Operating Manual ...