2. Setting Up for Phased Array Measurement
Page 26
PHASOR XS Operating Manual
Step 1:
Activate the BASE Submenu (located in the UT
Menu) by pressing
below it. Function can also be ac-
cessed via the SYNCH Submenu located in the SCAN
Menu.
Step 2:
Press
next to the function titled DISPLAY
START. Continue pressing
or turn the Function Knob
to select from:
IP START—Display start corresponds to the initial
pulse
IF START—Display start corresponds to the IF-Gate
trigger
MATERIAL—Display start corresponds to the posi-
tion at which sound enters the test material
NOTE:
To select a DISPLAY START other than MATE-
RIAL, the IF Gate must be turned ON. Initial IF Gate
positioning is related to the Z parameter, which must
not be set to zero (accessed via the WEDGE GEO Sub-
menu, found in the PROBE Menu). The Z parameter
relates to the measured distance from the probe to the
test piece through the delay material. The IF Gate is
initially positioned based on this value. Switching on
the IF Gate activates the IF condition that controls the
A- and B-Gate starting point.
2.7 Pulser and Receiver Settings for
Phased Array Operation
2.7.1 Setting the Pulser Voltage Level
(UT-PULSER-VOLTAGE)
The relative energy with which the pulser fires is adjusted
by changing the VOLTAGE setting. To set the pulser volt-
age level:
Step 1:
Activate the PULSER Submenu (located in the
UT Menu) by pressing
below it. Functions will appear
down the left side of the display screen.
Step 2:
Press
next to the function titled VOLTAGE.
Note that the pulser VOLTAGE level can be set between
50 and 150 volts by turning the Function Knob.
Step 3:
The pulser VOLTAGE will be set to the value last
displayed.
FIGURE 2-10—Changing the number of displayed LEGs modifies the displayed sound-path range. In this case,
approximately 2.5 legs are displayed here.
Summary of Contents for Phasor 16/16
Page 6: ...Contents Page 4 PHASOR XS Operating Manual ...
Page 16: ...1 Phased Array Operation Keypad Menus and Displays Page 14 PHASOR XS Operating Manual ...
Page 24: ...2 Setting Up for Phased Array Measurement Page 22 PHASOR XS Operating Manual ...
Page 86: ...5 Making Conventional Measurements Page 84 PHASOR XS Operating Manual ...
Page 92: ...6 Storing Data Sets and Generating Reports Page 90 PHASOR XS Operating Manual ...
Page 96: ...7 Technical Specifications Page 94 PHASOR XS Operating Manual ...