4. Conventional Operation: Menu System, Keypad, and Displays
PHASOR XS Operating Manual
Page 61
press
next to RECORD again. The value in the func-
tion box will now read “S-REF2?”.
Step 6:
Apply couplant and couple the probe to the thicker
calibration standard. Press
next to the selection
titled A START. Turn the function knob (this will shift the
starting point of the A-Gate) until the A-Gate lies over the
echo corresponding to the thicker standard’s thickness
(
Figure 4-12
).
Step 7:
Press
next to the selection titled RE-
CORD. The value in the function box will change from
“S-REF2 ?” to “OFF”. The instrument will automatically
calculate the material’s velocity and the probe delay.
NOTE:
The instrument’s AUTOCAL Function applies
only to conventional operation,
not
to phased-array
operation.
Checking Calibration Results
Following the calibration procedure, the calculated acous-
tical velocity and probe delay are displayed. To view these
calculated values:
Step 1:
Access the HOME Menu.
Step 2:
You’ll find these selections:
•
PROBE DELAY
—Adjustment made as a result of
the AUTOCAL (zeroing) procedure. This represents
the time delay caused by sound-wave travel through
a probe’s membrane, wear plate, or delay line.
•
VELOCITY
—Display’s the calculated velocity after
calibration. (Note that the material type is listed as
custom)
4.10
Using the Calibration Reminder
Alarm
The instrument incorporates a timed alarm feature that
causes an icon to appear on user input intervals between
0.5 and 4.0 hours. To use the alarm:
Step 1:
Access the CALALARM Submenu (located in the
CONFIG Menu).
FIGURE 4-12—Auto Calibration Procedures
Summary of Contents for Phasor 16/16
Page 6: ...Contents Page 4 PHASOR XS Operating Manual ...
Page 16: ...1 Phased Array Operation Keypad Menus and Displays Page 14 PHASOR XS Operating Manual ...
Page 24: ...2 Setting Up for Phased Array Measurement Page 22 PHASOR XS Operating Manual ...
Page 86: ...5 Making Conventional Measurements Page 84 PHASOR XS Operating Manual ...
Page 92: ...6 Storing Data Sets and Generating Reports Page 90 PHASOR XS Operating Manual ...
Page 96: ...7 Technical Specifications Page 94 PHASOR XS Operating Manual ...