7-10
MAC™ 1600
2028451-182B
Stress Testing
For more information on making these adjustments, refer to
8. When the exerecise phase is complete, press the
Recovery
key to
begin the recovery phase of the test.
NOTE
When using an ergometer with remote start, you do not need to
press the
Recovery
key because the recovery phase begins
automatically at the end of the last stage. You can, however,
press the
Recovery
key to begin the recovery phase before the
last stage ends.
On supported treadmills, the belt begins to slow and the grade drops
to 0%. On supported ergometers, the load begins to lighten. On
unsupported treadmills and ergometers, these adjustments must be
made manually.
Continue to monitor the patient and record the ECG until the device
stops.
9. When the recovery phase is over, press the
Test End
key.
The menu options at the bottom of the screen change to
Confirm Test
End
and
Continue Test
. Do one of the following:
To return to the test, press the
F2
key (
Continue Test
).
The previous menu options return. Continue to record the ECG
as needed. When you are done, repeat this step.
To stop the test, press the
F1
key (Confirm Test End).
The menu options change. Continue to step 10.
10. Do any of the following, as necessary.
Press the
F1
key (
Next Patient
) to test another patient.
You are warned that testing another patient will discard the
results of the current test. Do one of the following:
Press the
F5
key (
No
) to cancel the change in patients and
return to the current test.
You can either print the current test report or change the
report formats.
Press the
F6
key (
Yes
) to lose the current test results and test
a new patient.
Return to step 3 for the next patient.
Press the
F2
key (
) to print the test’s report.
The report prints with the selected format options.
Press the
F3
key (
Report Format
) to modify the report format.
The
Report Format
window opens. Select the options you want to
include in the report and press the
F6
key (
Save
). You can now
print the test’s report.
Summary of Contents for MAC 1600
Page 8: ...vi MAC 1600 2028451 182B...
Page 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Page 24: ...1 16 MAC 1600 2028451 182B Introduction...
Page 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Page 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Page 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Page 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Page 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Page 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Page 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Page 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Page 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Page 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Page 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Page 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Page 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Page 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Page 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Page 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...
Page 161: ......