Stress Testing
2028451-182B
MAC™ 1600
7-9
ergometers with remote start. Deviations for specific devices are noted
where appropriate.
DANGER
PATIENT INJURY — When on a moving treadmill, a
patient could fall and sustain an injury.
To minimize the possibility of a falling caused by the belt’s
sudden movement, have the patient step onto the belt
only after it begins moving.
When conducting stress tests on a supported treadmill,
press the
Stop TM
button twice to immediately stop the
belt in the case of an emergency (for example, if the
patient stumbles or falls while the belt is moving).
1. Prepare the patient as described in Chapter 3.
2. From the
Main Menu
, press
F3
to select the
Stress ECG
option.
The
Enter Patient Data
window opens.
3. Enter patient data as described in Chapter 4.
4. Adjust the stress options as necessary:
This includes the speed and gain, finite residual filter, pacemaker
enhancement, printer leads, test protocol, report format, and target
heart rate. For more information on setting these options, refer to
5. Record a preliminary ECG.
This may be a seated, standing, supine, or hyperventilating ECG,
depending on the requirements of the selected protocol.
6. Begin the pretest phase.
a. Have the patient get on the device.
b. Press the
Pretest
key.
c. Allow the patient to warm up before beginning the exercise phase
of the test.
NOTE
On supported treadmills, press
Start TM
to start the belt.
7. When the patient is ready to begin the stress test, press the
Exercise
key.
During the test, you can use the stress keys to hold the current stage,
enter blood pressure, add a comment, change the displayed leads,
and toggle the finite residual filter. With supported devices, the
stress keys can also be used to adjust the device’s speed, grade, or
load. With unsupported devices, the device must be adjusted
manually at the device itself.
Summary of Contents for MAC 1600
Page 8: ...vi MAC 1600 2028451 182B...
Page 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Page 24: ...1 16 MAC 1600 2028451 182B Introduction...
Page 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Page 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Page 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Page 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Page 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Page 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Page 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Page 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Page 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Page 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Page 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Page 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Page 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Page 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Page 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Page 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...
Page 161: ......