System Configuration
2028451-182B
MAC™ 1600
9-17
Most treadmill and ergometer protocols consist of three pre-defined
phases
:
Pretest
,
Exercise
, and
Recovery
. Each phase can include multiple
stages, which define the parameters of the test. The parameters differ
slightly depending on the device, as seen in the following table.
Protocols for Master’s Step devices consist of only
Stage
and
Stage Time
.
You can modify the pre-defined protocols to create custom protocols. Use
the following instructions to create a custom protocol.
1. From the MAC 1600
Main Menu
, press
F5
(
System Configuration
) >
F4
(
Stress ECG Setup
).
The
Stress ECG Setup
window opens.
2. Press
F4
(
Page Down
).
The second page opens.
Ergometers
WHO
WHO50
WHO75
HOLLMANN
BAL
STD.FRANCE
MODWHO
CONCONI
Master’s Step
SINGLE
DOUBLE
TRIPLE
Parameter
Treadmill
Ergometer
Comment
Stage
Y
Y
The stage name.
Stage Time
Y
Y
The stage duration, in minutes.
Speed
Y
N
The treadmill speed in kilometers or
miles per hour, depending on the
Country Setup.
Grade [%]
Y
N
The percentage of increase in the
treadmill’s elevation.
Basic Load [W]
N
Y
The load at which the ergometer
operates, in watts.
Store Median First
Y
Y
The interval at which the first median
reading is stored.
Store Median
Repeat
Y
Y
The interval at which a subsequent
median reading is stored.
BP First
Y
Y
The interval at which the first blood
pressure reading is stored.
BP Repeat
Y
Y
The interval at which subsequent
blood pressure readings are stored.
Device
Protocols
Summary of Contents for MAC 1600
Page 8: ...vi MAC 1600 2028451 182B...
Page 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Page 24: ...1 16 MAC 1600 2028451 182B Introduction...
Page 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Page 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Page 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Page 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Page 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Page 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Page 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Page 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Page 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Page 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Page 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Page 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Page 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Page 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Page 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Page 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...