Troubleshooting
2028451-182B
MAC™ 1600
A-3
System Will Not Power Up
If the system will not power up, do the following:
Verify the unit is turned on.
If it is not, turn the unit on. Refer to
page 2-11 for instructions.
Verify the battery is installed and charged.
Refer to
on page A-7 for instructions on verifying
whether the battery is installed and charged.
Refer to
on page 10-12 for instructions on
installing the battery.
Verify the unit is connected to an AC power outlet.
Refer to
“Connecting the AC Power Adapter”
on page 2-8 for
instructions.
Verify the equipment is receiving power from the outlet.
If the unit is receiving power, the Power LED will be lit.
ECG Data Contains Noise
If the acquired ECG data displays unacceptable noise levels, do the
following:
Check the patient’s position.
The patient should remain motionless during the acquisition of a
resting ECG.
Use the
Hookup Advisor
indicator to help determine the cause of the
noise.
For more information, refer to
Verify the electrodes are placed properly.
Refer to
on page 3-3 for information on
proper electrode placement.
Verify the electrodes have been applied correctly.
Perspiration, excessive hair, lotions, and dead skin cells must be
removed from the electrode site.
Refer to
on page 3-2 more information.
Check for defective or expired electrodes.
Replace the electrodes if there are any questions about their
effectiveness.
Check for defective, broken, or disconnected leadwires.
Replace the leadwires if there are any questions about their
effectiveness. Refer to
Consider using filters,
ADS
, and
FRF
to help eliminate or reduce
ECG noise.
For more information, refer to
on page 6-3, or
Summary of Contents for MAC 1600
Page 8: ...vi MAC 1600 2028451 182B...
Page 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Page 24: ...1 16 MAC 1600 2028451 182B Introduction...
Page 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Page 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Page 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Page 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Page 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Page 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Page 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Page 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Page 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Page 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Page 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Page 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Page 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Page 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Page 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Page 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...
Page 161: ......