System Configuration
2028451-182B
MAC™ 1600
9-9
Preview before Analysis
Determines waveform preview options. Values include:
No
Waveforms are never previewed.
Always
Waveforms are always previewed.
Yellow electrodes
Waveforms are previewed when the Hookup Advisor
indicator shows a yellow or red electrode.
Red electrodes
Waveforms are previewed when the Hookup Advisor
indicator shows a red electrode.
For additional information, refer to
Reanalysis
Enables/disables the reanalysis feature, which allows you
to adjust the following ECG measurements:
P Duration
PR Interval
QRS Duration
QT Interval
Available only if Audit Trail is disabled and one of the
following options is activated: ME12, MEHR, MI12, or
MIHR. For more information on activating options, refer to
For more information on the reanalysis feature, refer to
QTC Calculation
Determines which formula will be used to correct QT
calculations. Options are:
Bazett
Framingham
Fridericia
In all formulas, HR = Heart Rate. Bazett is available only if
the MEHR or MIHR option is activated. Framingham and
Fridericia are available only if the ME12 or MI12 option is
activated.
Field
Comment
QTc
QT
HR
60
--------
=
QTc
QT
154 1
60
HR
--------
–
⎝
⎠
⎛
⎞
+
=
QTc
QT
HR
60
--------
3
=
Summary of Contents for MAC 1600
Page 8: ...vi MAC 1600 2028451 182B...
Page 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Page 24: ...1 16 MAC 1600 2028451 182B Introduction...
Page 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Page 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Page 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Page 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Page 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Page 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Page 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Page 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Page 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Page 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Page 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Page 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Page 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Page 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Page 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Page 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...
Page 161: ......