Preparing the Patient
Stress ECG Electrodes
Standard 12 Lead Placement
To acquire a standard 12 lead ECG, use the placement shown in the
following illustration.
061A
AHA
Label
IEC Label
Electrode Placement
A
V1 red
C1 red
Fourth intercostal space at the right sternal border.
B
V2 yellow
C2 yellow
Fourth intercostal space at the left sternal border.
C
V3 green
C3 green
Midway between location B and D.
D
V4 blue
C4 brown
Mid-clavicular line in the fifth intercostal space.
E
V5 orange
C5 black
Anterior axillary line on the same horizontal level as
D.
F
V6 purple
C6 purple
Mid-axillary line on the same horizontal level as D
and E.
G
LA black
L yellow
Under left clavicle at the furthest point from the
sternum.
H
LL red
F green
On waistline, above left thigh.
Summary of Contents for MAC 1600
Page 8: ...vi MAC 1600 2028451 182B...
Page 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Page 24: ...1 16 MAC 1600 2028451 182B Introduction...
Page 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Page 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Page 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Page 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Page 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Page 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Page 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Page 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Page 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Page 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Page 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Page 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Page 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Page 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Page 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Page 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...
Page 161: ......