Recording a Resting ECG
2028451-182B
MAC™ 1600
5-5
F3
Gain
Changes the magnitude of the ECG signal on the
display or in the report. Measurement is in millimeter
per millivolt (mm/mV) and includes the following
options:
5 mm/mV
10 mm/mV
20 mm/mV
40 mm/mV
2.5 mm/mV
Automatic
The larger the selected measurement, the larger the
waveform appears. Only the appearance of the
waveform changes; signal strength is not affected.
NOTE
If Automatic is selected, the system calculates the
best gain based on the peak-to-peak amplitudes of
all displayed leads and the selected display
format.
F4
Filter
Eliminates noise in the waveform by restricting which
frequencies are included. Frequencies are measured in
Hertz (Hz) and include the following options:
20 Hz
40 Hz
100 Hz
150 Hz
Selecting a frequency eliminates signals that exceed
that frequency. The smaller the frequency selected, the
more signal is filtered out. For example, a filter of 40 Hz
displays only signals at 40 Hz or less; signals greater
than 40 Hz will be ignored.
CAUTION
INACCURATE READINGS — Using the filter can
result in a cleaner waveform, but selecting a frequency
that is too low could alter the waveform’s morphology,
resulting in an inaccurate reading.
To avoid this, use the filter only to eliminate excessive
noise and use the highest frequency that provides a
readable waveform.
F5
Pace Enhance
Improves the readability of pacemaker ECGs. Options
are On and Off.
F Key
Option
Description
Summary of Contents for MAC 1600
Page 8: ...vi MAC 1600 2028451 182B...
Page 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Page 24: ...1 16 MAC 1600 2028451 182B Introduction...
Page 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Page 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Page 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Page 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Page 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Page 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Page 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Page 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Page 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Page 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Page 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Page 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Page 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Page 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Page 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Page 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...
Page 161: ......