Managing Internal Storage
2028451-182B
MAC™ 1600
8-3
Importing Records
In addition to saving ECGs recorded with the MAC 1600 device, you can
also import ECG records to internal storage from the following sources:
Secure Digital (SD) cards,
Other MAC devices connected via serial port or modem,
CardioSoft systems connected via serial port or modem, or
MUSE systems connected via modem.
No additional set up is required to import from an SD card. However, you
must do the following to import data via serial port or modem:
Purchase and activate the appropriate communications option.
For more information refer to
Configure the system’s data communication settings.
For more information, refer to
Use the following instructions to import a record into internal storage.
1. From the
Main Menu
, press
F4
(
File Manager
).
The
File Manager
window opens.
2. Press
F3
(
Import
).
The function keys change.
051A
3. Select the appropriate import source:
To import ECGs from an SD card, insert the SD card and press
F1
(
SD Card
).
A list of the available ECGs on the card opens. Proceed to step 4.
To import ECGs via serial port, press
F2
(
Serial
).
The serial port opens. The device waits for the external device to
transmit records.
To import ECGs via modem, press
F3
(
Modem
).
The modem initializes. The device waits for the external device to
transmit records.
4. Select the records to be imported from the SD card.
Summary of Contents for MAC 1600
Page 8: ...vi MAC 1600 2028451 182B...
Page 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Page 24: ...1 16 MAC 1600 2028451 182B Introduction...
Page 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Page 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Page 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Page 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Page 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Page 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Page 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Page 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Page 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Page 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Page 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Page 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Page 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Page 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Page 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Page 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...
Page 161: ......