5-8
MAC™ 1600
2028451-182B
Recording a Resting ECG
If you press the
Rhythm
key after pressing the
Stop
key, the new report
will either begin printing immediately on the current sheet of paper or
advance to a new page, depending on the setting of the
Start rhythm
report on a new page
field on the
Resting ECG Setup
window. Refer to
on page 9-6 for details.
ECG Reanalysis
You can reanalyze ECGs if both of the following conditions have been
met:
Either the
Measurement and 12SL Interpretation
system option or
the
HEART
system option is enabled, and
Reanalysis
is selected on the
Resting ECG Setup
window.
Reanalysis allows you to modify the fiducial points on acquired
waveforms. The details of what modifications can be made depends on
which system option is enabled:
If the
Measurement and 12SL Interpretation
option is enabled,
reanalysis allows you to modify
Global Measurements
.
If the
HEART
option is enabled, reanalysis allows you to modify both
Global Measurements
and
T wave Dispersion
.
Reanalyzing an ECG
Use the following procedure to reanalyze a resting ECG.
For additional information, refer to
1. After acquiring an ECG, press the
F5
key (
Reanalyze
) to run the
Reanalyze
function.
For instructions on acquiring a resting ECG, refer to
2. Press the
F3
key to select the
Edit Mode
.
For more information, refer to
3. Review the waveforms to determine the accuracy of the system-
selected fiducial points.
For a better view of individual waveforms, use the
Leads
key to
toggle through the waveforms.
4. After you have analyzed the waveforms, use the following procedure
to adjust the fiducial points:
a. Press the
F1
key to toggle through the fiducial points.
The selected point changes size and is highlighted green.
Summary of Contents for MAC 1600
Page 8: ...vi MAC 1600 2028451 182B...
Page 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Page 24: ...1 16 MAC 1600 2028451 182B Introduction...
Page 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Page 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Page 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Page 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Page 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Page 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Page 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Page 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Page 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Page 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Page 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Page 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Page 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Page 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Page 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Page 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...
Page 161: ......