5-4
MAC™ 1600
2028451-182B
Recording a Resting ECG
7. When the waveforms are configured, press the
ECG
key to begin the
acquisition.
A progress bar indicates the percentage of the data acquired. When
the acquisition is complete, one of two things will occur, depending
on the setting of the
Preview Before Analysis
option on the
Resting
ECG Setup
window.
If the
Preview Before Analysis
option is enabled, a preview of the
10 second ECG is shown on the display. Proceed to step 8.
If the
Preview Before Analysis
option is not enabled, the ECG
data will be analyzed and printed after it has been acquired. Skip
to step 9.
8. While reviewing the preview, do one of the following.
To discard the reading and begin over, press
F3
(
Cancel
) and
repeat from step 4.
To accept the reading, press
F4
(
Continue
).
The menu options change to allow you to manage the acquisition.
Proceed to step 9.
9. Use the options to change patients, to print a copy, or to save,
transmit, or reanalyze the data.
For more information on each option, refer to
ECG Options
The MAC 1600 provides several options for configuring an ECG. The
options, presented as option keys across the bottom of the display, are
listed in the following table.
F Key
Option
Description
F1
Patient Data
Opens the patient data entry window.
F2
Speed
Changes the speed of the waveform on the display and
printout. Changing the speed also changes the speed
the wiper bar moves across the display.
Measurement is in millimeter per second (mm/s) and
includes the following options:
25 mm/s
50 mm/s
12.5 mm/s - 5 mm/s
12.5 mm/s
When the option includes two speeds (12.5 mm/s - 5
mm/s), the first speed is for the display and the second
speed is for the printout.
Summary of Contents for MAC 1600
Page 8: ...vi MAC 1600 2028451 182B...
Page 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Page 24: ...1 16 MAC 1600 2028451 182B Introduction...
Page 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Page 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Page 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Page 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Page 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Page 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Page 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Page 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Page 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Page 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Page 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Page 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Page 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Page 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Page 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Page 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...
Page 161: ......