![GE MAC 1600 Operator'S Manual Download Page 117](http://html1.mh-extra.com/html/ge/mac-1600/mac-1600_operators-manual_4251181117.webp)
System Configuration
2028451-182B
MAC™ 1600
9-25
Date of Birth
Determines whether date of birth will be available when
entering patient data.
Age
Determines whether age will be available when entering
patient data.
Height
Determines whether height will be available when entering
patient data.
Weight
Determines whether weight will be available when entering
patient data.
Gender
Determines whether gender will be available when
entering patient data.
Race
Determines whether race will be available when entering
patient data.
Phone Number
Determines whether phone number will be available when
entering patient data.
Pacemaker
Determines whether pacemaker will be available when
entering patient data.
Enable Patient ID Check
Determines whether additional checks will be performed to
ensure that the patient ID meets the requirements of the
national patient ID used in Scandinavian countries. If this
field is set, you must select the appropriate Patient ID
Type.
Patient ID Type
Available only in the Enable Patient ID Check field is set.
Determines which type of ID will be used and, therefore,
which checks to perform. Options are:
Swedish Patient ID
Danish Patient ID
Norwegian Patient ID
When a patient ID is entered, the system will verify its
format, extract the patient’s gender and date of birth, and
populate those fields if they have been enabled.
Patient ID Length (3-30)
Defines the maximum length of the patient ID within the
range of 3 to 30 characters.
Available only if the Enable Patient ID Check field is
cleared.
Sort Patient List by
Determines the field by which the patient list is sorted.
Options are:
Patient ID
Secondary ID
Patient Name
Field
Comment
Summary of Contents for MAC 1600
Page 8: ...vi MAC 1600 2028451 182B...
Page 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Page 24: ...1 16 MAC 1600 2028451 182B Introduction...
Page 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Page 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Page 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Page 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Page 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Page 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Page 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Page 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Page 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Page 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Page 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Page 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Page 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Page 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Page 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Page 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...
Page 161: ......