Arrhythmia Mode Recording
Introduction
The Arrhythmia mode is part of the basic MAC 1600 system. It allows
you to manually generate an arrhythmia printout in a table format, an
episode format, or a summary format.
The interface of the Arrhythmia mode is identical to the interface for the
Resting ECG mode. For more information on the interface, refer to
on page 5-2. In addition to the same waveform options
(speed, gain, filter, pace enhance, and patient data) as the Resting ECG
mode, arrhythmia mode also offers an anti-drifting system (ADS) that
helps reduce baseline shift.
Arrhythmia Mode
This section describes the process for recording an arrhythmia report,
the waveform options, and the printing options.
Printing an Arrhythmia Report
Use the following steps to record an arrhythmia report.
1. Prepare the patient as described in Chapter 3.
2. From the MAC 1600
Main Menu
, press the
F2
key (
Arrhythmia)
.
The
Enter Patient Data
window opens.
3. Enter the patient data as described in Chapter 4.
4. Adjust the gain, speed, filter, anti-drift system, and pacemaker
detection as necessary.
on page 6-3 for details.
5. When the settings have been adjusted as required, press the
F1
key
(
Start Recording
) to begin the arrhythmia report.
6. After you have recorded an adequate amount of information, press
the
F1
key (
Stop Recording
).
Two new options become available:
Confirm Stop
and
Continue
Recording
.
7. Do one of the following:
If additional information needs to be recorded, press the
F5
key
(
Continue Recording
).
This returns to the recording mode. Repeat from step 6.
If you have determined enough information has been recorded,
press the
F2
key (
Confirm Stop
).
Report options become available.
Summary of Contents for MAC 1600
Page 8: ...vi MAC 1600 2028451 182B...
Page 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Page 24: ...1 16 MAC 1600 2028451 182B Introduction...
Page 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Page 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Page 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Page 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Page 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Page 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Page 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Page 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Page 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Page 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Page 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Page 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Page 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Page 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Page 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Page 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...
Page 161: ......