FDA Acoustic Output Data
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
6–54
546L
Mode
MI
TIS/TIB
Scan
TIS
Non-
Scan
TIB
Non-
Scan
TIC
I
SPTA.3
I
SPPA.3
B-Mode
1.13
0.94
–
–
1.32
105.9
808.6
M-Mode
1.13
–
0.11
0.38
0.16
149.6
808.6
B/M-Mode
1.13
0.63
0.11
0.38
1.03
220.2
808.6
Pulsed Wave Doppler
1.26
–
0.39
1.25
0.54
529.2
832.8
B/Pulsed Wave Doppler
1.26
0.63
0.39
1.25
1.41
5599.8
832.8
B/Color Flow Doppler
1.42
2.88
–
–
3.94
351.5
808.6
B/Color Flow Doppler/
Pulsed Wave Doppler
1.42
1.75
0.20
0.62
2.68
475.6
832.8
M-Mode/Color M-Mode
1.13
–
0.52
1.64
0.72
575.4
808.6
Color Flow with M Doppler
1.13
–
0.52
1.64
0.72
575.4
808.6
Table 6–42. 546L Maximum Acoustic Output Levels
739L
Mode
MI
TIS/TIB
Scan
TIS
Non-
Scan
TIB
Non-
Scan
TIC
I
SPTA.3
I
SPPA.3
B-Mode
1.39
0.91
–
–
1.04
87.1
1117.2
M-Mode
1.39
–
< 0.1
0.28
0.10
139.1
1117.2
B/M-Mode
1.39
0.61
< 0.1
0.28
0.80
197.2
1117.2
Pulsed Wave Doppler
1.67
–
0.36
1.53
0.64
560.6
817.2
B/Pulsed Wave Doppler
1.67
0.61
0.36
1.53
1.33
618.7
1117.2
B/Color Flow Doppler
1.39
3.16
–
–
3.85
417.2
1117.2
B/Color Flow Doppler/
Pulsed Wave Doppler
1.67
1.88
0.18
0.77
2.59
518.0
1117.2
M-Mode/Color M-Mode
1.39
–
1.04
2.00
1.28
540.0
1117.2
Color Flow with M Doppler
1.39
–
1.04
2.00
1.28
540.0
1117.2
Table 6–43. 739L Maximum Acoustic Output Levels
Summary of Contents for Logiq 500