FDA Acoustic Output Data
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
6–51
C551
Mode
MI
TIS/TIB
Scan
TIS
Non-Scan
TIB
Non-Scan
TIC
I
SPTA.3
B-Mode
0.76
0.64
–
–
0.92
51.0
M-Mode
0.76
–
0.26
0.30
0.32
59.6
B/M-Mode
0.76
0.43
0.26
0.30
0.93
93.6
Pulsed Wave Doppler
0.98
–
0.54
2.38
1.83
656.9
B/Pulsed Wave Doppler
0.98
0.43
0.54
2.38
2.45
690.9
B/Color Flow Doppler
1.18
1.25
–
–
1.38
111.9
B/Color Flow Doppler/
Pulsed Wave Doppler
1.18
0.84
0.27
1.19
1.91
401.4
M-Mode/Color M-Mode
0.76
–
0.37
1.40
0.68
641.7
Color Flow with M Doppler
0.76
–
0.37
1.40
0.68
641.7
Table 6–36. C551 Maximum Acoustic Output Levels
C721
Mode
MI
TIS/TIB
Scan
TIS
Non-Scan
TIB
Non-Scan
TIC
I
SPTA.3
B-Mode
0.46
0.35
–
–
0.34
10.3
M-Mode
0.35
–
< 0.1
< 0.1
< 0.1
15.6
B/M-Mode
0.46
0.23
< 0.1
< 0.1
0.26
22.4
Pulsed Wave Doppler
0.42
–
0.31
0.24
0.28
91.8
B/Pulsed Wave Doppler
0.46
0.23
0.31
0.24
0.50
98.7
B/Color Flow Doppler
0.46
0.85
–
–
0.96
43.8
B/Color Flow Doppler/
Pulsed Wave Doppler
0.46
0.54
0.15
0.12
0.73
71.2
M-Mode/Color M-Mode
0.30
–
0.62
0.52
0.73
87.5
Color Flow with M Doppler
0.30
–
0.62
0.52
0.73
87.5
Table 6–37. C721 Maximum Acoustic Output Levels