FDA Acoustic Output Data
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
6–56
LD
Mode
MI
TIS/TIB
Scan
TIS
Non-Scan
TIB
Non-Scan
TIC
I
SPTA.3
B-Mode
0.63
0.21
–
–
0.49
4.4
M-Mode
0.63
–
<0.1
<0.1
<0.1
26.1
B/M-Mode
0.63
0.14
<0.1
<0.1
0.38
29.1
Pulsed Wave Doppler
0.58
–
0.33
2.32
1.70
680.6
B/Pulsed Wave Doppler
0.63
0.14
0.33
2.32
2.02
683.6
B/Color Flow Doppler
0.90
1.35
–
–
1.37
82.6
B/Color Flow Doppler/
Pulsed Wave Doppler
0.90
0.74
0.16
1.16
1.69
383.1
M-Mode/Color M-Mode
0.79
–
0.29
2.13
1.23
655.5
Color Flow with M Doppler
0.79
–
0.29
2.13
1.23
655.5
Table 6–46. LD Maximum Acoustic Output
P509
Mode
MI
TIS/TIB
Scan
TIS
Non-Scan
TIB
Non-Scan
TIC
I
SPTA.3
B-Mode
0.88
0.25
–
–
0.26
287.1
M-Mode
0.88
–
< 0.1
0.18
< 0.1
43.9
B/M-Mode
0.88
0.16
< 0.1
0.18
0.26
235.3
Pulsed Wave Doppler
0.81
–
0.56
1.63
0.66
405.9
B/Pulsed Wave Doppler
0.88
0.16
0.56
1.63
0.83
597.4
B/Color Flow Doppler
0.88
0.63
–
–
0.72
470.7
B/Color Flow Doppler/
Pulsed Wave Doppler
0.88
0.40
0.28
0.82
0.77
534.0
M-Mode/Color M-Mode
0.88
–
0.48
1.56
0.54
455.4
Color Flow with M Doppler
0.88
–
0.48
1.56
0.54
455.4
CWD
< 0.1
–
1.33
1.73
2.49
232.8
Table 6–47. P509 Maximum Acoustic Output Levels