Advanced Cardiac Calculations
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
4–91
Doppler Analysis – Aortic Valve (D-AV) (cont’d)
Calc Mnemonic
Calc Name
Input Measurements
Formula
AV-A
Aortic Valve Area by
cont’Eq
one distance
two velocities
AV-A[mm or cm]=
LVOT-A x PFV-LV/PFV-AV
LVOTA
Left Ventricular Outflow
Tract Area at Systole
LVOTs of PLAX
LVOTA[cm^2]=
π
x (LVOTs)^2 /4
AccT
Flow Acceleration Time
two time intervals
AccT[msec or sec]=
t<PFV-AV>-ts<ET>
DecT
Flow Deceleration Time
two time intervals
DecT[msec or sec]=
te<ET>-t<PFV-AV>
PkFVI
Peak Flow Velocity Intergal
one velocity,
one time interval
PkFVI[mm or cm]=
1.14 x PFV–AV x ET/
20+.03
MeanFV
Mean Flow Velocity
flow velocities,
one time interval
MeanFV[cm/s or m/s]=FVI/
t<FVI>
QP:QS
QP:QS Ratio
two CO
QP:QS=CO(PV)/
CO(AV)
MaxPG
Maximum Pressure
Gradient
two velocities
MaxPG[mmHg]=
4 x (PFV-AV^2)
MeanPG
Mean Pressure Gradient
flow velocities
MeanPG[mmHg]=
n
4 x
Σ
(V^2/n)
i=1
SV
Stroke Volume
flow velocities, one area
(by ellipse or trace or
circle)
AV–A is calculated.
SV[ml]=FVI–AV x AV–A
else
SV[ml]=FVI–AV x (
π
/4) X
(AOs)^2
SI
Stroke Volume Index
body surface area and SV
SI[ml/m^2]=
SV/BSA
CO
Cardiac Output
flow velocities, one area
(by ellipse, trace or circle)
and one time
interval
CO[ml]=
SVxHR/1000
CI
Cardiac Index
body surface area and CO
CI[ml/min/m^2]=
CO/BSA
Table 4–25. Aortic Valve (D-AV) (cont’d)