Advanced Cardiac Calculations
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
4–88
Doppler Analysis – Mitral Valve (D-MV)
F
O
R
s
e
a
d
P
V
p
2
2
1
Figure 4–76. D-MV Measurements
Calc Mnemonic
Calc Name
Input Measurements
Formula
MVOA
Mitral Valve Orifice Area at
Diastole
one area (by ellipse, trace
or circle)
MVOA=a1[cm ^2]
PFVMV
Peak Flow Velocity at the
Mitral Valve
one velocity
PFV–MV=v1[cm/s or m/s]
PFVLA
Peak Flow Velocity at Left
Atrium
one velocity
PFV–LA=v1[cm/s or m/s]
ET
Ejection Time
one time interval
ET=t1[ms or sec]
FVI
Flow Velocity Integral
flow velocities
FVI[mm or cm]=
Vdt
PHT
Pressure Half Time
one time interval
PHT=t1[ms orsec]
HR
Heart Rate (beats/minute)
one 2 beat time
interval
HR[BPM]=120 [sec]/ 2
beat time [sec]
Table 4–24. Mitral Valve (D-MV)
Summary of Contents for Logiq 500
Page 1: ......