5-26
Issue 01, 04/2005
Krautkramer USM 35X
Operation
Recording a reference echo
Before using the measurement of dB difference, you
have to first record a reference echo.
A
Attention:
When recording a reference echo, an already stored
reference echo is overwritten after a corresponding
warning.
– Peak the reference echo according to the test
specification.
– Position the A gate over the reference echo using the
function
aSTART
.
– Select the function
REFECHO
.
– Turn the right-hand rotary knob upward in order to
store the echo in gate A as a reference echo.
– If necessary, confirm the warning message in order
to overwrite a stored reference echo.
The reference echo is now recorded. An inverted
R
is
shown in the measurement line.
Deleting a reference echo
You can delete stored reference echoes.
– Select the function
REFECHO
.
– Turn the right-hand rotary knob downward in order to
delete the reference echo.
– If necessary, confirm the warning message in order
to delete the stored reference echo.
Measurement of dB difference (function group REF)
5-26
Issue 01, 04/2005
Krautkramer USM 35X
Operation
Recording a reference echo
Before using the measurement of dB difference, you
have to first record a reference echo.
A
Attention:
When recording a reference echo, an already stored
reference echo is overwritten after a corresponding
warning.
– Peak the reference echo according to the test
specification.
– Position the A gate over the reference echo using the
function
aSTART
.
– Select the function
REFECHO
.
– Turn the right-hand rotary knob upward in order to
store the echo in gate A as a reference echo.
– If necessary, confirm the warning message in order
to overwrite a stored reference echo.
The reference echo is now recorded. An inverted
R
is
shown in the measurement line.
Deleting a reference echo
You can delete stored reference echoes.
– Select the function
REFECHO
.
– Turn the right-hand rotary knob downward in order to
delete the reference echo.
– If necessary, confirm the warning message in order
to delete the stored reference echo.
Measurement of dB difference (function group REF)
Summary of Contents for Krautkramer USM 35X
Page 8: ...0 8 Issue 01 04 2005 Krautkramer USM 35X ...
Page 9: ...Krautkramer USM 35X Issue 01 04 2005 1 1 Introduction 1 Introduction 1 ...
Page 20: ...1 12 Issue 01 04 2005 Krautkramer USM 35X ...
Page 27: ...Krautkramer USM 35X Issue 01 04 2005 3 1 Initial start up 3 Initial start up 3 ...
Page 35: ...Krautkramer USM 35X Issue 01 04 2005 4 1 Principles of operation 4 Principles of operation 4 ...
Page 49: ...Krautkramer USM 35X Issue 01 04 2005 5 1 Operation 5 Operation 5 ...
Page 130: ...5 82 Issue 01 04 2005 Krautkramer USM 35X ...
Page 131: ...Krautkramer USM 35X Issue 01 04 2005 6 1 Documentation 6 Documentation 6 ...
Page 135: ...Krautkramer USM 35X Issue 01 04 2005 7 1 Maintenance and care 7 Maintenance and care 7 ...
Page 148: ...7 14 Issue 01 04 2005 Krautkramer USM 35X ...
Page 177: ...Krautkramer USM 35X Issue 01 04 2005 9 1 Appendix 9 Appendix 9 ...
Page 190: ...9 14 Issue 01 04 2005 Krautkramer USM 35X ...
Page 191: ...Krautkramer USM 35X Issue 01 04 2005 10 1 Changes 10 Changes 10 ...
Page 193: ...Krautkramer USM 35X Issue 01 04 2005 11 1 Index 11 Index 11 ...