5-8
Issue 01, 04/2005
Krautkramer USM 35X
Operation
Adjusting the display range (function group BASE)
If the display should for example start from the surface of
the test object, the value in
D-DELAY
must be set to 0.
• Coarse adjustment
–10 mm ... 1024 mm/–0.3" ... 40" in even steps
• Fine adjustment
up to 99.9 mm/9.999" in steps of 0.01 mm/0.001"
up to 1024 mm/10" in steps of 0.1 mm/0.001"
– Select the function
D-DELAY
.
– If required, toggle between coarse and fine adjust-
ment.
– Adjust the value for the display starting point by
means of the right-hand rotary knob.
P-DELAY (Probe delay)
Every probe has a delay line between the transducer
element and the coupling face. This means that the
initial pulse must first pass through this delay line be-
fore the sound wave can enter the test object. You can
compensate for this influence of the delay line in the
function
P-DELAY
.
H
Note:
If the value for
P-DELAY
is not known, read the section
Calibrating the USM 35X, chapter 5.7, in order to deter-
mine this value.
– Select the function
P-DELAY
.
– Adjust the value for the probe delay by means of the
right-hand rotary knob.
5-8
Issue 01, 04/2005
Krautkramer USM 35X
Operation
Adjusting the display range (function group BASE)
If the display should for example start from the surface of
the test object, the value in
D-DELAY
must be set to 0.
• Coarse adjustment
–10 mm ... 1024 mm/–0.3" ... 40" in even steps
• Fine adjustment
up to 99.9 mm/9.999" in steps of 0.01 mm/0.001"
up to 1024 mm/10" in steps of 0.1 mm/0.001"
– Select the function
D-DELAY
.
– If required, toggle between coarse and fine adjust-
ment.
– Adjust the value for the display starting point by
means of the right-hand rotary knob.
P-DELAY (Probe delay)
Every probe has a delay line between the transducer
element and the coupling face. This means that the
initial pulse must first pass through this delay line be-
fore the sound wave can enter the test object. You can
compensate for this influence of the delay line in the
function
P-DELAY
.
H
Note:
If the value for
P-DELAY
is not known, read the section
Calibrating the USM 35X, chapter 5.7, in order to deter-
mine this value.
– Select the function
P-DELAY
.
– Adjust the value for the probe delay by means of the
right-hand rotary knob.
Summary of Contents for Krautkramer USM 35X
Page 8: ...0 8 Issue 01 04 2005 Krautkramer USM 35X ...
Page 9: ...Krautkramer USM 35X Issue 01 04 2005 1 1 Introduction 1 Introduction 1 ...
Page 20: ...1 12 Issue 01 04 2005 Krautkramer USM 35X ...
Page 27: ...Krautkramer USM 35X Issue 01 04 2005 3 1 Initial start up 3 Initial start up 3 ...
Page 35: ...Krautkramer USM 35X Issue 01 04 2005 4 1 Principles of operation 4 Principles of operation 4 ...
Page 49: ...Krautkramer USM 35X Issue 01 04 2005 5 1 Operation 5 Operation 5 ...
Page 130: ...5 82 Issue 01 04 2005 Krautkramer USM 35X ...
Page 131: ...Krautkramer USM 35X Issue 01 04 2005 6 1 Documentation 6 Documentation 6 ...
Page 135: ...Krautkramer USM 35X Issue 01 04 2005 7 1 Maintenance and care 7 Maintenance and care 7 ...
Page 148: ...7 14 Issue 01 04 2005 Krautkramer USM 35X ...
Page 177: ...Krautkramer USM 35X Issue 01 04 2005 9 1 Appendix 9 Appendix 9 ...
Page 190: ...9 14 Issue 01 04 2005 Krautkramer USM 35X ...
Page 191: ...Krautkramer USM 35X Issue 01 04 2005 10 1 Changes 10 Changes 10 ...
Page 193: ...Krautkramer USM 35X Issue 01 04 2005 11 1 Index 11 Index 11 ...