8-10
Issue 01, 04/2005
Krautkramer USM 35X
Interfaces and Peripherals
•
Entry
of a new value or state of a function using the
command structure:
<ESC> <COMMAND> <SPACE> <VALUE> <RE TURN>
All values are entered or transmitted by the USM 35X
without a comma or a point. The resolution of the func-
tion should therefore be observed with all values. The
resolution of a function applies to the entire value
range of that function.
A
resolution of 0.01
means:
The USM 35X transmits the value of a function multi-
plied by the factor of 100. The entry of a value must be
done multiplied by the factor of 100.
Examples
• Setting of the display delay to 72.39 mm:
<ESC>dd 7239 <RETURN>
• Setting of the display width to 72.3 mm:
<ESC>dw 7230 <RETURN>
Remote control
• Setting of the display width to 192 mm:
<ESC>dw 19200 <RETURN>
A
resolution of 0.1
means:
The USM 35X transmits the value of a function multi-
plied by the factor of 10. The entry of a value must be
done multiplied by the factor of 10.
Example:
• Setting of the gain to 51.5 dB
<ESC>db 515 <RETURN>
A
resolution of 1
means:
The USM 35X transmits the value of a function without
multiplication. The entry of a value must be done with-
out any multiplication.
Example:
• Setting the response threshold of gate a to 41 %:
<ESC>at 41 <RETURN>
8-10
Issue 01, 04/2005
Krautkramer USM 35X
Interfaces and Peripherals
•
Entry
of a new value or state of a function using the
command structure:
<ESC> <COMMAND> <SPACE> <VALUE> <RE TURN>
All values are entered or transmitted by the USM 35X
without a comma or a point. The resolution of the func-
tion should therefore be observed with all values. The
resolution of a function applies to the entire value
range of that function.
A
resolution of 0.01
means:
The USM 35X transmits the value of a function multi-
plied by the factor of 100. The entry of a value must be
done multiplied by the factor of 100.
Examples
• Setting of the display delay to 72.39 mm:
<ESC>dd 7239 <RETURN>
• Setting of the display width to 72.3 mm:
<ESC>dw 7230 <RETURN>
Remote control
• Setting of the display width to 192 mm:
<ESC>dw 19200 <RETURN>
A
resolution of 0.1
means:
The USM 35X transmits the value of a function multi-
plied by the factor of 10. The entry of a value must be
done multiplied by the factor of 10.
Example:
• Setting of the gain to 51.5 dB
<ESC>db 515 <RETURN>
A
resolution of 1
means:
The USM 35X transmits the value of a function without
multiplication. The entry of a value must be done with-
out any multiplication.
Example:
• Setting the response threshold of gate a to 41 %:
<ESC>at 41 <RETURN>
Summary of Contents for Krautkramer USM 35X
Page 8: ...0 8 Issue 01 04 2005 Krautkramer USM 35X ...
Page 9: ...Krautkramer USM 35X Issue 01 04 2005 1 1 Introduction 1 Introduction 1 ...
Page 20: ...1 12 Issue 01 04 2005 Krautkramer USM 35X ...
Page 27: ...Krautkramer USM 35X Issue 01 04 2005 3 1 Initial start up 3 Initial start up 3 ...
Page 35: ...Krautkramer USM 35X Issue 01 04 2005 4 1 Principles of operation 4 Principles of operation 4 ...
Page 49: ...Krautkramer USM 35X Issue 01 04 2005 5 1 Operation 5 Operation 5 ...
Page 130: ...5 82 Issue 01 04 2005 Krautkramer USM 35X ...
Page 131: ...Krautkramer USM 35X Issue 01 04 2005 6 1 Documentation 6 Documentation 6 ...
Page 135: ...Krautkramer USM 35X Issue 01 04 2005 7 1 Maintenance and care 7 Maintenance and care 7 ...
Page 148: ...7 14 Issue 01 04 2005 Krautkramer USM 35X ...
Page 177: ...Krautkramer USM 35X Issue 01 04 2005 9 1 Appendix 9 Appendix 9 ...
Page 190: ...9 14 Issue 01 04 2005 Krautkramer USM 35X ...
Page 191: ...Krautkramer USM 35X Issue 01 04 2005 10 1 Changes 10 Changes 10 ...
Page 193: ...Krautkramer USM 35X Issue 01 04 2005 11 1 Index 11 Index 11 ...