![GE Krautkramer USM 35X Operator'S Manual Download Page 63](http://html.mh-extra.com/html/ge/krautkramer-usm-35x/krautkramer-usm-35x_operators-manual_128397063.webp)
Krautkramer USM 35X
Issue 01, 04/2005
5-15
Operation
used, i.e. when function parameters are changed.
Pos-
sible alarms occuring during instrument operation
(setting of functions) are to be ignored.
Display of gates
To make the assignment easier, the gates are displayed
in different colors. You cannot vary the colors of gates
because they are fixed as follows:
• Gate A – red
• Gate B – green
• Gate C – blue
aLOGIC/bLOGIC
(Evaluation logic of the gates)
This function allows you to choose the method for trig-
gering the gate alarm. The alarm is output to the LED
A
on the front panel of the USM 35X. There are four set-
ting options available:
Setting the gates (function groups aGAT and bGAT)
•
off
– Evaluation logic off
The alarm and measurement capability are switched
off. The gate is not visible.
•
pos
– Coincidence
The alarm (LED A) is on if the preset response
threshold of the gate is exceeded within the dis-
played range.
•
neg
– Anticoincidence
The alarm (LED A) is on if the preset response
threshold of the gate is not reached within the
displayed range.
•
a trig
– Triggering by interface echo
When using gate A as echo-start gate (setting of the
evaluation logic for gate B)
– Select the function
aLOGIC
or
bLOGIC
.
– Set the required alarm logic by means of the right-
hand rotary knob.
H
Note:
The alarm and measurement function of the gates is
only active within the display range.
Krautkramer USM 35X
Issue 01, 04/2005
5-15
Operation
used, i.e. when function parameters are changed.
Pos-
sible alarms occuring during instrument operation
(setting of functions) are to be ignored.
Display of gates
To make the assignment easier, the gates are displayed
in different colors. You cannot vary the colors of gates
because they are fixed as follows:
• Gate A – red
• Gate B – green
• Gate C – blue
aLOGIC/bLOGIC
(Evaluation logic of the gates)
This function allows you to choose the method for trig-
gering the gate alarm. The alarm is output to the LED
A
on the front panel of the USM 35X. There are four set-
ting options available:
Setting the gates (function groups aGAT and bGAT)
•
off
– Evaluation logic off
The alarm and measurement capability are switched
off. The gate is not visible.
•
pos
– Coincidence
The alarm (LED A) is on if the preset response
threshold of the gate is exceeded within the dis-
played range.
•
neg
– Anticoincidence
The alarm (LED A) is on if the preset response
threshold of the gate is not reached within the
displayed range.
•
a trig
– Triggering by interface echo
When using gate A as echo-start gate (setting of the
evaluation logic for gate B)
– Select the function
aLOGIC
or
bLOGIC
.
– Set the required alarm logic by means of the right-
hand rotary knob.
H
Note:
The alarm and measurement function of the gates is
only active within the display range.
Summary of Contents for Krautkramer USM 35X
Page 8: ...0 8 Issue 01 04 2005 Krautkramer USM 35X ...
Page 9: ...Krautkramer USM 35X Issue 01 04 2005 1 1 Introduction 1 Introduction 1 ...
Page 20: ...1 12 Issue 01 04 2005 Krautkramer USM 35X ...
Page 27: ...Krautkramer USM 35X Issue 01 04 2005 3 1 Initial start up 3 Initial start up 3 ...
Page 35: ...Krautkramer USM 35X Issue 01 04 2005 4 1 Principles of operation 4 Principles of operation 4 ...
Page 49: ...Krautkramer USM 35X Issue 01 04 2005 5 1 Operation 5 Operation 5 ...
Page 130: ...5 82 Issue 01 04 2005 Krautkramer USM 35X ...
Page 131: ...Krautkramer USM 35X Issue 01 04 2005 6 1 Documentation 6 Documentation 6 ...
Page 135: ...Krautkramer USM 35X Issue 01 04 2005 7 1 Maintenance and care 7 Maintenance and care 7 ...
Page 148: ...7 14 Issue 01 04 2005 Krautkramer USM 35X ...
Page 177: ...Krautkramer USM 35X Issue 01 04 2005 9 1 Appendix 9 Appendix 9 ...
Page 190: ...9 14 Issue 01 04 2005 Krautkramer USM 35X ...
Page 191: ...Krautkramer USM 35X Issue 01 04 2005 10 1 Changes 10 Changes 10 ...
Page 193: ...Krautkramer USM 35X Issue 01 04 2005 11 1 Index 11 Index 11 ...