5-16
Issue 01, 04/2005
Krautkramer USM 35X
Operation
aSTART/bSTART
(Starting points of the gates)
You can fix the starting point of the gates A or B
within
the adjustment range of 0 ... 9999 mm/250".
– Select the function
aSTART
or
bSTART
.
– Uuse the right-hand rotary knob to adjust the required
setting.
aWIDTH/bWIDTH (Width of the gates)
You can determine the gate width within the range of
0.2 ... 9999 mm/0.008 ... 250".
– Select the function
aWIDTH
or
bWIDTH
.
– Use the right-hand rotary knob to adjust the required
value.
aTHRSH/bTHRSH (Response and mea-
surement threshold of the gates)
You can determine the threshold value of the gates within
the range of 10 to 90 % screen height for triggering the
LED alarm if this value is exceeded or not reached, de-
pending on the setting of the
aLOGIC/bLOGIC
function.
In the RF mode, the threshold can be additionally set
from –90 % to –10 %.
– Select the function
aTHRSH
or
bTHRSH
.
– Set the required value using the right-hand rotary
knob.
Setting the gates (function groups aGAT and bGAT)
5-16
Issue 01, 04/2005
Krautkramer USM 35X
Operation
aSTART/bSTART
(Starting points of the gates)
You can fix the starting point of the gates A or B
within
the adjustment range of 0 ... 9999 mm/250".
– Select the function
aSTART
or
bSTART
.
– Uuse the right-hand rotary knob to adjust the required
setting.
aWIDTH/bWIDTH (Width of the gates)
You can determine the gate width within the range of
0.2 ... 9999 mm/0.008 ... 250".
– Select the function
aWIDTH
or
bWIDTH
.
– Use the right-hand rotary knob to adjust the required
value.
aTHRSH/bTHRSH (Response and mea-
surement threshold of the gates)
You can determine the threshold value of the gates within
the range of 10 to 90 % screen height for triggering the
LED alarm if this value is exceeded or not reached, de-
pending on the setting of the
aLOGIC/bLOGIC
function.
In the RF mode, the threshold can be additionally set
from –90 % to –10 %.
– Select the function
aTHRSH
or
bTHRSH
.
– Set the required value using the right-hand rotary
knob.
Setting the gates (function groups aGAT and bGAT)
Summary of Contents for Krautkramer USM 35X
Page 8: ...0 8 Issue 01 04 2005 Krautkramer USM 35X ...
Page 9: ...Krautkramer USM 35X Issue 01 04 2005 1 1 Introduction 1 Introduction 1 ...
Page 20: ...1 12 Issue 01 04 2005 Krautkramer USM 35X ...
Page 27: ...Krautkramer USM 35X Issue 01 04 2005 3 1 Initial start up 3 Initial start up 3 ...
Page 35: ...Krautkramer USM 35X Issue 01 04 2005 4 1 Principles of operation 4 Principles of operation 4 ...
Page 49: ...Krautkramer USM 35X Issue 01 04 2005 5 1 Operation 5 Operation 5 ...
Page 130: ...5 82 Issue 01 04 2005 Krautkramer USM 35X ...
Page 131: ...Krautkramer USM 35X Issue 01 04 2005 6 1 Documentation 6 Documentation 6 ...
Page 135: ...Krautkramer USM 35X Issue 01 04 2005 7 1 Maintenance and care 7 Maintenance and care 7 ...
Page 148: ...7 14 Issue 01 04 2005 Krautkramer USM 35X ...
Page 177: ...Krautkramer USM 35X Issue 01 04 2005 9 1 Appendix 9 Appendix 9 ...
Page 190: ...9 14 Issue 01 04 2005 Krautkramer USM 35X ...
Page 191: ...Krautkramer USM 35X Issue 01 04 2005 10 1 Changes 10 Changes 10 ...
Page 193: ...Krautkramer USM 35X Issue 01 04 2005 11 1 Index 11 Index 11 ...