Krautkramer USM 35X
Issue 01, 04/2005
5-19
Operation
Calibrating the USM 35X
Case B: With unknown material velocity
Use the semiautomatic calibration function of the
USM 35X via the function group
CAL
for this calibration
case.
The distances between 2 calibration echoes must be
entered as default data. The USM 35X will then carry
out a plausibility check, calculate the material velocity
and the probe delay, and automatically set the parame-
ters.
Calibration process
– Set the required display range in
RANGE
(function
group
BASE
). The two calibraion echoes selected
must be displayed on the screen. Set the range so
that the second calibration echo is located on the
right edge of the screen.
– Select the function group
CAL
.
– Enter the distances of the two calibation echoes in
S-REF1
and
S-REF2
.
– Position the gate (function
aSTART
) on the first
calibration echo.
– Press
to record the first calibration echo.
– The recording of the first calibration echo is con-
firmed by the message “
Echo is recorded
”, and the
function
CAL
indicates the value
1
.
– Move the gate to the second calibration echo.
– Press
to record the second calibration echo.
Krautkramer USM 35X
Issue 01, 04/2005
5-19
Operation
Calibrating the USM 35X
Case B: With unknown material velocity
Use the semiautomatic calibration function of the
USM 35X via the function group
CAL
for this calibration
case.
The distances between 2 calibration echoes must be
entered as default data. The USM 35X will then carry
out a plausibility check, calculate the material velocity
and the probe delay, and automatically set the parame-
ters.
Calibration process
– Set the required display range in
RANGE
(function
group
BASE
). The two calibraion echoes selected
must be displayed on the screen. Set the range so
that the second calibration echo is located on the
right edge of the screen.
– Select the function group
CAL
.
– Enter the distances of the two calibation echoes in
S-REF1
and
S-REF2
.
– Position the gate (function
aSTART
) on the first
calibration echo.
– Press
to record the first calibration echo.
– The recording of the first calibration echo is con-
firmed by the message “
Echo is recorded
”, and the
function
CAL
indicates the value
1
.
– Move the gate to the second calibration echo.
– Press
to record the second calibration echo.
Summary of Contents for Krautkramer USM 35X
Page 8: ...0 8 Issue 01 04 2005 Krautkramer USM 35X ...
Page 9: ...Krautkramer USM 35X Issue 01 04 2005 1 1 Introduction 1 Introduction 1 ...
Page 20: ...1 12 Issue 01 04 2005 Krautkramer USM 35X ...
Page 27: ...Krautkramer USM 35X Issue 01 04 2005 3 1 Initial start up 3 Initial start up 3 ...
Page 35: ...Krautkramer USM 35X Issue 01 04 2005 4 1 Principles of operation 4 Principles of operation 4 ...
Page 49: ...Krautkramer USM 35X Issue 01 04 2005 5 1 Operation 5 Operation 5 ...
Page 130: ...5 82 Issue 01 04 2005 Krautkramer USM 35X ...
Page 131: ...Krautkramer USM 35X Issue 01 04 2005 6 1 Documentation 6 Documentation 6 ...
Page 135: ...Krautkramer USM 35X Issue 01 04 2005 7 1 Maintenance and care 7 Maintenance and care 7 ...
Page 148: ...7 14 Issue 01 04 2005 Krautkramer USM 35X ...
Page 177: ...Krautkramer USM 35X Issue 01 04 2005 9 1 Appendix 9 Appendix 9 ...
Page 190: ...9 14 Issue 01 04 2005 Krautkramer USM 35X ...
Page 191: ...Krautkramer USM 35X Issue 01 04 2005 10 1 Changes 10 Changes 10 ...
Page 193: ...Krautkramer USM 35X Issue 01 04 2005 11 1 Index 11 Index 11 ...