5-76
Issue 01, 04/2005
Krautkramer USM 35X
Operation
– If required, confirm by pressing
again.
After the successful recording of the reference echo, a
highlighted
R
will be displayed in the measurement line.
– Set the function
DGSMOD>
to
on
, to switch on DGS
curve.
Taking the general DGS diagram as a basis, the instru-
ment calculates the required test sensitivity for display-
ing the 3 mm curve with its maximum at 80 % screen
height, and sets this value. The current gain is set to 0
during this. In the case of subsequent gain variations,
the curve is automatically adapted.
Evaluation according to the DGS method (only USM 35X S)
Evaluation of reflectors
Every echo situated within the gate can be immediately
evaluated:
The measurement line has been configured in such a
way that the sound path
S
, the echo height in
dB
referred to the DGS curve
, the equivalent reflector
size (
ERS
) of the flaw echo and the calibration range
are displayed.
The measured value to be zoomed in the A-scan is
chosen by means of the function
S-DISP
in the func-
tion group
MEAS.
ERS
was chosen in the above exam-
ple. (Also refer to
Configuring the measurement line,
p. 5-49.)
5-76
Issue 01, 04/2005
Krautkramer USM 35X
Operation
– If required, confirm by pressing
again.
After the successful recording of the reference echo, a
highlighted
R
will be displayed in the measurement line.
– Set the function
DGSMOD>
to
on
, to switch on DGS
curve.
Taking the general DGS diagram as a basis, the instru-
ment calculates the required test sensitivity for display-
ing the 3 mm curve with its maximum at 80 % screen
height, and sets this value. The current gain is set to 0
during this. In the case of subsequent gain variations,
the curve is automatically adapted.
Evaluation according to the DGS method (only USM 35X S)
Evaluation of reflectors
Every echo situated within the gate can be immediately
evaluated:
The measurement line has been configured in such a
way that the sound path
S
, the echo height in
dB
referred to the DGS curve
, the equivalent reflector
size (
ERS
) of the flaw echo and the calibration range
are displayed.
The measured value to be zoomed in the A-scan is
chosen by means of the function
S-DISP
in the func-
tion group
MEAS.
ERS
was chosen in the above exam-
ple. (Also refer to
Configuring the measurement line,
p. 5-49.)
Summary of Contents for Krautkramer USM 35X
Page 8: ...0 8 Issue 01 04 2005 Krautkramer USM 35X ...
Page 9: ...Krautkramer USM 35X Issue 01 04 2005 1 1 Introduction 1 Introduction 1 ...
Page 20: ...1 12 Issue 01 04 2005 Krautkramer USM 35X ...
Page 27: ...Krautkramer USM 35X Issue 01 04 2005 3 1 Initial start up 3 Initial start up 3 ...
Page 35: ...Krautkramer USM 35X Issue 01 04 2005 4 1 Principles of operation 4 Principles of operation 4 ...
Page 49: ...Krautkramer USM 35X Issue 01 04 2005 5 1 Operation 5 Operation 5 ...
Page 130: ...5 82 Issue 01 04 2005 Krautkramer USM 35X ...
Page 131: ...Krautkramer USM 35X Issue 01 04 2005 6 1 Documentation 6 Documentation 6 ...
Page 135: ...Krautkramer USM 35X Issue 01 04 2005 7 1 Maintenance and care 7 Maintenance and care 7 ...
Page 148: ...7 14 Issue 01 04 2005 Krautkramer USM 35X ...
Page 177: ...Krautkramer USM 35X Issue 01 04 2005 9 1 Appendix 9 Appendix 9 ...
Page 190: ...9 14 Issue 01 04 2005 Krautkramer USM 35X ...
Page 191: ...Krautkramer USM 35X Issue 01 04 2005 10 1 Changes 10 Changes 10 ...
Page 193: ...Krautkramer USM 35X Issue 01 04 2005 11 1 Index 11 Index 11 ...