![GE Hydran 201i Instruction Manual Download Page 129](http://html.mh-extra.com/html/ge/hydran-201i/hydran-201i_instruction-manual_128390129.webp)
Verification of Operation Prior to Installation
Part 17997
Rev. 6, March 2011
6-15
8.
Modify the
Analog Out Mode
by pushing
CHANGE
.
9.
When asked to enter the password #1 (
1253
), use the
UP
and
ENTER
keys.
10.
Push
DOWN
three times and
ENTER
to force the output to
0 %
.
11.
Allow about one minute for the output to stabilize.
12.
Verify the ammeter reads 4.00 ± 0.10 mA.
13.
Push
CHANGE
,
UP
and
ENTER
to force the output to
50 %
.
14.
Allow about one minute for the output to stabilize.
15.
Verify the ammeter reads 12.00 ± 0.20 mA.
16.
Push
CHANGE
,
UP
and
ENTER
to force the output to
100 %
.
17.
Allow about one minute for the output to stabilize.
18.
Verify the ammeter reads 20.00 ± 0.40 mA.
19.
Push
CHANGE
,
UP
and
ENTER
to return the output to
Normal
.
20.
When all is done, push
END
.
21.
Disconnect the ammeter.
6.2.9
Verifying the RS-232 Communication (Facultative)
This procedure requires the use of the Hydran Host software. Please refer to the
Hydran Host Software Manual
.
1.
Using the supplied RS-232 cable, connect the Hydran 201T
i
to a host computer
through their DB-9 connector. Refer to Table D-7 on page D-4 for connection details.
2.
Access the top of the extended menu by pushing
ENTER
twice (
History
appears).
Note: Once accessed, the extended menu times out after 30 minutes of keypad inactivity.
Summary of Contents for Hydran 201i
Page 1: ...GE Energy Services Digital Energy Hydran Instruction Manual 201i System ...
Page 6: ...Instruction Manual vi Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 8: ...Instruction Manual viii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 22: ...Instruction Manual xxii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 36: ...Chapter 1 Introduction 1 10 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 114: ...Chapter 5 Before Installation 5 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 143: ...Installation Part 17997 Rev 6 March 2011 6 29 Figure 6 9 Installing the CPU Module ...
Page 236: ...Chapter 12 Troubleshooting 12 20 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 244: ...Chapter 13 Maintenance 13 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 254: ...Appendix B Mechanical Drawings B 2 Rev 6 March 2011 Part 17997 Figure B 3 1 5 In Adaptor ...
Page 255: ...Hydran 201Ti Intelligent Transmitter Part 17997 Rev 6 March 2011 B 3 Figure B 4 1 In Adaptor ...
Page 284: ...Appendix G Unit Conversions G 2 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 298: ...Appendix I Glossary I 12 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 299: ......