Part 17997
Rev. 6, March 2011
A-1
Appendix A
Technical Specifications for the Hydran 201
i
System
A.1
HYDRAN 201T
I
INTELLIGENT TRANSMITTER
GENERAL
Description
Continuous, on-line, intelligent gas-in-oil transmitter
Components
Sensor and electronic modules housed in cylindrical enclosure
Gas Response
Hydrogen (H
2
), carbon monoxide (CO), acetylene (C
2
H
2
), ethylene (C
2
H
4
)
Medium
Mineral, insulating oil for transformers
Application
Transformer monitoring; specifically, detection of incipient faults in oil-filled,
electrical equipment
ANALYTICAL PERFORMANCE
Principle
Gas-permeable membrane and combustible gas detector
Sampling Method
Flooded port with 1-in NPT male threads
Measurement Range
0–2,000 ppm (volume/volume, H
2
equivalent); other ranges are available
Accuracy
For 0–2,000 ppm range: ± 10 % of reading ± 25 ppm (H
2
equivalent)
Relative Sensitivity
• H
2
: 100 % of concentration
• CO: 18 ± 3 % of concentration
• C
2
H
2
: 8 ± 2 % of concentration
• C
2
H
4
: 1.5 ± 0.5 % of concentration
Response Time
10 minutes (90 % of step change)
External Sampling
Port
Designed for glass syringe with Luer stop cock; closed with 5/32-in Allen
screw
ELECTRONIC UNIT
Hardware
Microprocessor; watchdog; clock
Software
Real-time operating system; menu-driven interface
Functions
• Gas level, hourly trend and daily trend readings
• Gas level, gas trends and fail alarms
• History logging
• Periodic sensor test
• Calibration, configuration and self-test
• Networking
• Remote control via Hydran 201C
i
Controller (optional modem)
• Remote, embedded software upgrading
Communications
User selection (one port at a time):
• RS-232 link (DB-9 connector) to connect host computer
• Supervisory link to connect H201C
i
Controller
Summary of Contents for Hydran 201i
Page 1: ...GE Energy Services Digital Energy Hydran Instruction Manual 201i System ...
Page 6: ...Instruction Manual vi Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 8: ...Instruction Manual viii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 22: ...Instruction Manual xxii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 36: ...Chapter 1 Introduction 1 10 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 114: ...Chapter 5 Before Installation 5 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 143: ...Installation Part 17997 Rev 6 March 2011 6 29 Figure 6 9 Installing the CPU Module ...
Page 236: ...Chapter 12 Troubleshooting 12 20 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 244: ...Chapter 13 Maintenance 13 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 254: ...Appendix B Mechanical Drawings B 2 Rev 6 March 2011 Part 17997 Figure B 3 1 5 In Adaptor ...
Page 255: ...Hydran 201Ti Intelligent Transmitter Part 17997 Rev 6 March 2011 B 3 Figure B 4 1 In Adaptor ...
Page 284: ...Appendix G Unit Conversions G 2 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 298: ...Appendix I Glossary I 12 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 299: ......