Chapter 2 • Hydran 201T
i
Intelligent Transmitter
2-2
Rev. 6, March 2011
Part 17997
• A DB-9 connector for local RS-232 serial communication with a host computer running
the Hydran Host software
• Serial communication with a Hydran 201C
i
Controller through a supervisory link
(communication cable between a H201T
i
and a H201C
i
)
• Full stand-alone capabilities when the host computer (running the Hydran Host software)
is not used
• A very wide temperature range for all-weather outdoor operation
• A thermal regulating system
• Terminal blocks to connect the following components:
– ac power supply
– Alarm contacts (gas High and High-High, and system fail alarms)
– Analog output
• A termination board to connect the H201C
i
and the H201T
i
with a supervisory link
2.2
ENCLOSURE
The Hydran 201T
i
’s enclosure (see Figure 1-1 on page 1-3) is 186 mm (7.3 in) in diameter
by 180 mm (7.125 in) in length; for details, see Figure B-1 on page B-1. It serves three
functions:
• To provide mechanical and weather protections for the Hydran 201 sensor and the
H201T
i
’s electronic circuits
• To maintain the H201 sensor and the H201T
i
electronics between 25 and 50 °C (77 and
122 °F) under extreme ambient and oil temperatures
• Using the dynamic oil sampling system, to stimulate oil movement in front of the H201
sensor to ensure representative sampling at all times
2.3
EXTERNAL COMPONENTS
Here is a description of the Hydran 201T
i
’s external components (see Figure 2-1 on
1.
Display Window
: The aluminum cover has a window to see the LCD display. The
display and the keypad are located on the CPU module (Section 2.4.4 on page 2-7).
2.
Connection Box
: It is mounted on the external side of the heating plate, under the
valve. It has a removable, watertight cover and two holes to install standard conduit
fittings (for watertight, rigid or flexible steel conduits).
Summary of Contents for Hydran 201i
Page 1: ...GE Energy Services Digital Energy Hydran Instruction Manual 201i System ...
Page 6: ...Instruction Manual vi Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 8: ...Instruction Manual viii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 22: ...Instruction Manual xxii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 36: ...Chapter 1 Introduction 1 10 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 114: ...Chapter 5 Before Installation 5 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 143: ...Installation Part 17997 Rev 6 March 2011 6 29 Figure 6 9 Installing the CPU Module ...
Page 236: ...Chapter 12 Troubleshooting 12 20 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 244: ...Chapter 13 Maintenance 13 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 254: ...Appendix B Mechanical Drawings B 2 Rev 6 March 2011 Part 17997 Figure B 3 1 5 In Adaptor ...
Page 255: ...Hydran 201Ti Intelligent Transmitter Part 17997 Rev 6 March 2011 B 3 Figure B 4 1 In Adaptor ...
Page 284: ...Appendix G Unit Conversions G 2 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 298: ...Appendix I Glossary I 12 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 299: ......