9
08/24/2016
6. Edition, ID No.: 37020GB-F
Operating Manual ERESCO MF4 Type Series
Copyright by GE Sensing & Inspection Technologies GmbH Bogenstrasse 41 D - 22926 Ahrensburg
Our broadly automated control units make monitoring during the exposure time
dispensable. The operator can therefore immediately increase the distance after
switching-on the high voltage. The effect of this measure is increased since after
switch-on (approx. 6 sec./100kV) the high voltage is slowly increased from zero to
the preselected value in the case of all
GE Sensing & Inspection Technologies
control
units.
9. Always remember to cover the rear side of the film with an additional lead plate
when taking shots. By doing so, you substantially reduce general radiation level and
additionally protect the film against stray radiation that reduces contrast.
10. Please always remember that so long as X-ray radiation is being generated, the yellow
signal lamp on the control unit lights or flashes or if necessary the warning lamp or
optionally the flash lamp works.
11. Never forget to remove the key of control unit and to secure the unit against unauthorised
users during working breaks.