2.1.1
Operation with optional warning flash lamp
Depending on the available power supply 230 V or 115 V, a warning flash lamp can be
optionally connected optionally to socket
X3
.
(Use the supplied eyebolt to remove the short-circuit plug)
NOTE:
With a connected warning flash lamp, an prewarning time of at least 2
seconds m u s t be set, otherwise the high voltage cannot be switched
on.
2.1.2
Operation with connected door contacts
When the safety door contacts are to be connected, use the supplied eyebolt to remove
the short circuit plug for the door contact from socket
X2
at the plug hutch.
NOTE:
If door contact 1 has been opened, door contact 2 should also be opened.
Both door contacts must be closed before the unit is ready for operation
again.
Short circuit plug with eyebolt
Eyebolt
Short circuit plug
Fig. 5 - Plug hutch of the digital control unit
ERESCO MF4 control (without cover, without handle)
X1
Mains connection
X2
Door contacts /
Potential-free
contacts
X3 Warning flash lamp
X4
Cooling circuit
X5
Radiating unit
X6
External Start/Stop
Fuses
F1
and
F2
Horn
Earthing Bold