Operating Manual ERESCO MF4 Type Series
Copyright by GE Sensing & Inspection Technologies GmbH Bogenstrasse 41 D - 22926 Ahrensburg
Press the
X-RAY ON button to start the warm-up program.
The X-ray tube is warmed up to the new high voltage value, as described in section 2.2.4.
2.8
Exposure Stop
The exposure can be stopped at any time by pressing the
X-RAY OFF
button. The high
voltage is reduced to zero before the high voltage supply is switched off.
In this case, the displayed actual-value of the exposure time remains at the current value,
and the exposure can be continued at any time by restarting. An exposure time error
cannot occur.
If a new exposure is to be carried out after the
manual
stop, the exposure time must be
reset as described under point 2.3.1.
If the high voltage is set higher than the warm-up voltage, the message 106: Warm-up
necessary!
appears on the display (also see section 2.7).