Operating Manual ERESCO MF4 Type Series
Copyright by GE Sensing & Inspection Technologies GmbH Bogenstrasse 41 D - 22926 Ahrensburg
1.8
Intended Use
The unit is exclusively designed for material testing or material radiation.
The unit should only be used for the described purpose in compliance with the safety
notes
according to Section 1 to 1.7.
Any other use of the unit is strictly prohibited.
Improper use or tampering with the unit may result in personal injuries or damage to the
unit.
1.6 Personal
Protective Clothing
Always wear the specified protective clothing at the
installation site!
1.7
Safety Measures at the Installation Site
Always use the shielding equipment available at site. Structural conditions can quite often
be used to enhance safety.
You should clearly demarcate and label control areas properly.
ATTENTION: Increased earth leakage current.
The earth leakage current of this X-ray equipment is higher than 3.5 mA.
According to EN 60204-1 an additional (separate) ground cable with the
at least same wire gauge (wire cross section) as the wires in the mains
cable must be connected from the earthing bolt at the control unit to a
ground terminal, that has a minimum wire gouge of 10 mm² Cu (AWG7).
X1
X1
X5
Radiating unit
Control unit
Connecting cable
Power supply cable
230 V / 115 V
min. 10 mm² Cu
additional ground