Recommended procedure for message 82, 94 and 111
The messages 82, 94 and 111 can occur when the radiating unit
· is only operated occasionally at the maximum high voltage.
· has not been operated for a longer period or is only operated occasionally.
· is operated for short operating cycles.
· is operated in extremely high or extremely low ambient temperatures.
To ensure normal operation, GE Sensing & Inspection Technologies recommends conditioning
the radiating unit in
power mode
. To achieve this, the following operating parameters are
required:
ERESCO 160 MF4-R
60 kV
10 mA
10 min
ERESCO 200 MF4-R
60 kV
10 mA
10 min
ERESCO 42 MF4
90 kV
10 mA
10 min
ERESCO 32 MF4-C
60 kV
10 mA
10 min
ERESCO 65 MF4
150 kV
6 mA
10 min
ERESCO 52 MF4-CL
100 kV
6 mA
10 min
ERESCO 280 MF4-R
100 kV
3.4 mA
10 min
ERESCO 300 MF4-R
150 kV
6 mA
10 min
After successful operation in power mode as described above, normal operation can be
resumed.
Note:
If after execution of the said procedure the radiating unit cannot be
operated at the desired high voltage, the power mode procedure can be
repeated up to three times. If this does not result in correct operation of
the radiating unit, please contact the Customer Service Department of
GE Sensing & Inspection Technologies.