43
08/24/2016
6. Edition, ID No.: 37020GB-F
Operating Manual ERESCO MF4 Type Series
Copyright by GE Sensing & Inspection Technologies GmbH Bogenstrasse 41 D - 22926 Ahrensburg
04 Prewarning
The current
prewarning time is displayed.
Press the
F5
key to exit the menu item without changes.
Press the
F2
key to change the function key arrangement and to display the current
prewarning time inversely:
This setting will be changed by entering another value via the numeric keypad or by pressing
the
or
keys.
Press the
OK key to confirm the new setting.
Enter
0 to switch off the prewarning time.
After entering 0 second, the
password 1483 must be entered to confirm the setting.
Password input is not required for prewarning time changes above 2 seconds. A prewarning
time setting of 1 second is not possible.
CAUTION:
The prewarning time is designed to protect personnel and should not be
set to 0!
NOTE:
With a connected warning flash lamp, an prewarning time of at least 2
seconds
m u s t be set, otherwise the high voltage cannot be switched on.
The prewarning time is factory set to 2 seconds.
Current prewarning time in
seconds.