Programming Site Data
1-35
March 2008
Selecting Error Handling
(cont.)
For the
AVE
mode, the
2-path error handling
option is intended for
applications where two sets of transducers are installed in the same
location in the same pipe to improve accuracy. With this function
enabled, the Model GS868 performs error handling only if both
channels are in error. If this function is disabled, error handling
occurs when either channel goes into error.
•
Press
[F1] = NO
to disable 2-path error handling, or
•
Press
[
F2] = YES
to enable 2-path error handling.
Specific responses of the display and the totalizer to the 2-path error
handling option are listed in Table 1-10 below.
The
User Program
automatically returns to the
Global I/O
prompt.
Table 1-10:
2-Path Error Response Options
Option
Display Response
Totalizer Response
NO
Displays the average of CH1
and CH2, regardless of the
error state of either channel.
Outputs the average of CH1
and CH2 totals, regardless
of the error state of either
channel.
YES
1. If one channel is in error,
the other channel’s value
is displayed as average.
2. If both channels are in
error, the last average
reading is held.
1. If one channel is in error,
totalizing continues.
2. If both channels are in
error, totalizing stops.
Summary of Contents for DigitalFlow GS868
Page 7: ...Chapter 1 ...
Page 59: ...Chapter 2 ...
Page 77: ...Chapter 3 ...
Page 90: ...Chapter 4 ...
Page 104: ...Chapter 5 ...
Page 110: ...Chapter 6 ...
Page 122: ...Appendix A ...
Page 131: ...Appendix B ...
Page 132: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2 ...
Page 136: ...Appendix C ...
Page 156: ...Appendix D ...