March 2008
4-12
Printing Data
Printing RTD Data
Use the RTDs submenu to send the numeric point value of a
connected RTD device to the RS232 port for display on a terminal or
RS232 printer. By viewing the point value of the meter at a given
temperature, you can calculate the points at the Set Temperature, and
the slope of the RTD input in points/degree. (See Chapter 1,
Calibration
, of the
Service Manual
for more details.)
1.
To access the
Print Menu
, press the
[PRNT]
key.
Note:
The
Print Menu
is not protected by the security feature and a
password is never required to access this menu.
2.
Press the [
←]
or [
→]
key to display the option bar shown, and
then press
[F3]
to select the
RTDs
submenu.
3.
At the
Dump RTD Data
prompt, press
[F1]
to leave the
menu
without dumping data, or
[F2]
to send the RTD data to the RS232
port.
After the selection has been made, the Model GS868 returns to the
standard data display screen and normal measurement taking. If you
have selected
YES
at the
Dump RTD Data
prompt, the meter continues
to send data until you reenter the
menu and select the
STOP
option.