March 2008
C-4
Programming with PanaView
Setting up Ethernet
Communications
If you have selected TCP/IP in step 6 on the previous page, the Setup
Communications window appears similar to Figure C-3 below.
Figure C-3: Setup Communications for TCP/IP
7.
Type in the desired
Name
and
Timeout
(in milliseconds).
8.
In the
IP Address
text box, enter the IP address. If the IP address is
unknown, run the
Device Discovery
software utility. All units found
will be identified by their MAC address and the assigned IP
address. In the
Port Number
box, enter 2101 (the default value).
9.
Click
[OK]
to complete data entry.
IMPORTANT:
If you are using Ethernet communications, be sure
the GS868 is set up with the default communications
parameters: 9600 baud, no parity, no handshaking, 8
data bits and 1 stop bit. The port number must match
the “Enable Raw TCP access using TCP Port”
assignment shown under the
Device Discovery
“TCP
Server Settings” menu (
Configuration>Serial
Ports>Port
).