Revision C
250cx Series Maternal/Fetal Monitor
3-17
2036947-001
Installation: Service Mode Screens
FM Remote Mark
This field configures the marker annotation that is printed on the strip chart paper
whenever a patient presses the button on the Remote Marker accessory. Use the
Trim Knob control to alternate between
On
and
Off
. The factory default setting is
Off
.
The
annotation is commonly used to record an “event”; while the
annotation is commonly used as an indication that the mother has perceived fetal
movement.
SpO
2
Scale
Two scale options are available for printing the MSpO
2
trends. The scale is printed
on the paper along with the trend.
Auto:
The trend plots on an expanded scale of 60–100% or
50–100%, depending on the paper.
1
0–100%:
This option configures the MSpO
2
trend to always plot at a fixed scale
of 0–100%.
Fetal Alert/Alarms
If the Spectra Alerts option is installed in your monitor, use this field to select
between the built-in FHR alarm features of the monitor and the Spectra Alerts
option.
Alarms:
The monitor generates alarms based on the limit settings provided on the
FHR setup screens.
Alerts:
Enables the Spectra Alerts feature to analyze heart rate and uterine
activity data to detect certain abnormal trends and alert the clinician.
Alert Suspend
When a care provider is at the patient’s bedside, it may be desirable to suspend the
audio component of alerts.
Off
(disabled)
:
users are unable to suspend audio alerts.
On
(enabled)
:
users can manually activate/de-activate the function.
Re-Alarm
This field adjusts the temporary silence period. An audio alarm is cancelled using
the
Alarm Silence
button. However, for MECG and MSpO
2
monitoring and during a
paper-load error condition, an alarm will be re-issued if the alarm state continues
after a specified amount of time.
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 6: ...CE CE ii 0086 ...
Page 14: ...viii 250cx Series Maternal Fetal Monitor Revision C 2036947 001 ...
Page 16: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036947 001 2 Equipment Overview ...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 55: ...3 1 250cx Series Maternal Fetal Monitor Revision C 2036947 001 3 Installation ...
Page 56: ...3 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 81: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036947 001 4 Maintenance ...
Page 82: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 141: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036947 001 5 Troubleshooting ...
Page 142: ...5 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 194: ...6 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 237: ...Revision C 250cx Series Maternal Fetal Monitor B 1 2036947 001 B Alarms Summary ...
Page 238: ...B 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Alarms Summary For your notes ...
Page 251: ......