5-10
250cx Series Maternal/Fetal Monitor
Revision C
2036947-001
Troubleshooting: Diagnostic Control Screen
1.
Loosen the four hex-head lock screws—two on each side. The right side screws
are labeled
B
in the figure on page
.
2.
To move the printhead forward on one side, back-off the corresponding captive
screw (turn counterclockwise) from its alignment block. To move the printhead
backward on one side, tighten the corresponding captive screw (turn
clockwise).
3.
After making the necessary adjustments, tighten the four hex-head lock screws
labeled
B
.
Thermal Printhead
The thermal printhead is a part of the printhead assembly. Refer to the parts list in
Chapter 6,
“Parts List, Drawings, and Replacement”
, of this manual for
ordering information.
Removing the Printhead
Use the following procedure for removing the 250cx Series Monitor printhead. As
long as you have a short screwdriver, you will not need to remove the strip chart
recorder from the monitor.
1.
Turn off the 250cx Series Monitor and disconnect the power cord from the
monitor.
2.
Remove the nine screws which secure the monitor top cover. Four screws are
located on the bottom of the monitor and five screws are located in the back.
3.
Remove the cover by sliding it toward the rear of the monitor.
4.
Disconnect the printhead harness cable from J3 on the Recorder Board.
Table 14. Printhead Adjustments
Adjustment
Description
A
Set screws loosened to perform a vertical
offset adjustment.
B
Hex-head screw loosened to perform a
horizontal offset adjustment.
C
Hex-head screw which determines the
vertical offset of a horizontal line. (See
.)
D
Captive hex-head screw which controls
the horizontal offset of a vertical line on
the top grid of the strip chart paper (heart
rate channel).
E
Captive hex-head screw which controls
the horizontal offset of a vertical line on
the bottom grid of the strip chart paper
(uterine pressure channel).
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 6: ...CE CE ii 0086 ...
Page 14: ...viii 250cx Series Maternal Fetal Monitor Revision C 2036947 001 ...
Page 16: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036947 001 2 Equipment Overview ...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 55: ...3 1 250cx Series Maternal Fetal Monitor Revision C 2036947 001 3 Installation ...
Page 56: ...3 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 81: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036947 001 4 Maintenance ...
Page 82: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 141: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036947 001 5 Troubleshooting ...
Page 142: ...5 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 194: ...6 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 237: ...Revision C 250cx Series Maternal Fetal Monitor B 1 2036947 001 B Alarms Summary ...
Page 238: ...B 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Alarms Summary For your notes ...
Page 251: ......